Precision Dimensional Measurement Specialist | ADMT
042-482-0621master@admt.krKoreanEnglish

RENISHAW AGILITY SERIES

AGILITY S 12129 R

Raised working volume for pallet loading

Renishaw AGILITY 5-axis CMM series
Measuring range1,200 × 1,200 × 900 mm
Length measurement errorAccuracy (MPEE):
1.2 + L/280 µm @ 20 °C
ConfigurationRaised working volume for pallet loading

MODEL SPECIFICATIONS

Detailed model data

Review measuring volume, accuracy, environmental limits and scanning performance.

Measuring range (X × Y × Z)1,200 × 1,200 × 900 mm
Accuracy (MPEE)1.2 + L/280 µm @ 20 °C
Metrology temperature15 °C to 40 °C
Operating temperature15 °C to 40 °C
Temperature fluctuationper hour 2 °C / per day 8 °C
Temperature gradientper metre 2 °C
Relative humidity20% ~ 70% RH (with air conditioning 80%)
Minimum air supplyMinimum 6.5 bar (REVO-2 fitted)
Maximum part weight1,000 kg
Y-axis configurationDual encoder / dual linear motor
Scanning errorMPE(THN) 2.1 µm
Scanning timeMPT(t) 9 seconds
Single-stylus form errorMPE(PFTU) 1.4 µm
Roundness form errorMZCI(RONt) 1.3 µm (with FeatureCal 0.8 µm)

L is the measured length in millimetres. MPE(E0)/MPE(E150) follows ISO 10360-2:2009; scanning and probing values follow the stated data-sheet test conditions.

MACHINE DIMENSIONS

Overall dimensions and full specifications

Review the dimensions, system requirements, environmental limits and accuracy performance required for installation planning.

AGILITY S 12129 R overall dimension drawing
AGILITY S 12129 R · overall dimensions and measuring range

System specification

Measuring range X
1,200 mm
Measuring range Y
1,200 mm
Measuring range Z
900 mm
Overall width W
2,235 mm
Overall length L
2,610 mm
Overall height H
3,550 mm
H1
550 mm
H2
211 mm
H3
44.5 mm
Zmax
1,220 mm
L2
100 mm
Machine mass
5,200 kg

Environmental conditions

Metrology temperature
15 °C ~ 40 °C
Temperature fluctuations
per hour 2 °C / per day 8 °C
Temperature gradient
per metre 2 °C
Operating temperature
15 °C ~ 40 °C
Relative humidity
20% ~ 70% RH (with air conditioning 80%)
Minimum air supply pressure
Minimum 6.5 bar (REVO-2 fitted)

Machine specification

Maximum axis velocity
300 mm/s
Maximum axis acceleration
800 mm/s²
Maximum vector velocity
520 mm/s
Maximum vector acceleration
1,400 mm/s²
Thread size
Bed M10 / inside wall M8
Maximum part weight
1,000 kg
Control system
UCC S5
Joystick options
MCU5-2 / MCU W-2
Y-axis configuration
Dual encoder / dual linear motor

Accuracy and measuring performance

MPE(E0) / MPE(E150)
1.2 + L/280 µm @ 20 °C
MPL(R0)
1.1 µm
MPE(THN)
2.1 µm
MPT(t)
9 seconds
MPE(PFTU)
1.4 µm
MPE(PFTj)
2.7 µm
MPE(PSTj)
1.0 µm
MPE(PLTj)
2.2 µm
MZCI(RONt)
1.3 µm (with FeatureCal 0.8 µm)
Axis measurement
Renishaw VIONiC readheads with ZeroMet and steel scale

Accuracy verification criteria

MPE(E0) / MPE(E150)Length measurement error and length measurement error with 150 mm offset, ISO 10360-2:2009

MPE(THN) / MPT(t)Scanning error on a datum sphere and scanning time, ISO 10360-4:2000

MPE(PFTU / PFTj / PSTj / PLTj)Single and multi-stylus probing performance, ISO 10360-5:2010

MZCI(RONt)Roundness form measuring error, ISO 12181

Renishaw REVO 5-axis high-speed scanning application
5-AXIS

REVO 5-axis measurement

Synchronous REVO head and CMM-axis motion captures complex geometry quickly while reducing structure-related dynamic errors.

Renishaw REVO tactile and non-contact multi-sensor family
MULTI-SENSOR

Multi-sensor capability

Configure tactile scanning, non-contact vision and surface-finish sensing to suit the inspection objective.

Renishaw MODUS 5-axis measurement programming interface
SOFTWARE

MODUS inspection programming

Connect CAD-based planning, DMIS programs, 5-axis path simulation and result analysis in one inspection workflow.

START A CONVERSATION

Have a measurement question?
Tell us what you need—we'll be glad to help.